Model 14 Oxygen Analyzer
5.1.1 CALIBRATION Menus
Any calibration should follow the procedure below:
1. Configure the calibration characteristics in the “PROGRAMMING” menu.
2. Perform the calibration via the “EXECUTION” menu.
5-2
An access code may be required if one has been programmed,
see page 5-14 for CODE Menu.
The Temp. Calib. menu will not appear when instrument is set for manual
temperature compensation.
m
DATE mm/dd/yy Date of the last calibration. The date programmed is not updated automatically.
S x.xxx nA/ppm Slope value
T x.x°C Drift between the theoretical temperature (sensor curve) T
h
and the temperature
measured T
m
:DT = T
h
- T
m
Parameters
CALIBRATION
CONC. CALIB. TEMP. CALIB.
PARAMETERS
Select
Select Select
CONC. CALIB.
TEMP. CALIB.
PARAMETERS
HISTORIC
ZERO
SLOPE
EXECUTION
DATE: 01/01/98
S: 5.40nA/ppm
Zero: 0.00nA
T: 0.0 ˚C
Enter
Enter
Enter