Broadley Technologies 14 Manual de instrucciones Pagina 47

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 46
Model 14 Oxygen Analyzer
Causes and solutions:
A) The membrane permeability has changed (coating).
Calibrate the analyzer and check if the
concentration is back to normal.
B) Electrolyte contamination.
Inspect probe components. Change the
electrolyte.
C) Leak of electrolyte.
Inspect probe components. Change the
electrolyte.
D) Error during calibration or incorrect calibration.
Calibrate again to check the parameters.
If the error is confirmed, check the calibration
current (too high, too low or unstable) and the
concentration in the air. Refer to the problems
described above.
E) The temperature calibration has not been
performed.
Check the temperature given by the trans-
mitter and calibrate it (see page 6-2). If
manual compensation has been selected,
confirm the value is correct.
F) The sample temperature or pressure is out of the
specifications.
Change the probe location or modify the
sample so that it meets the specifications.
Causes and solutions:
A) The probe current is zero.
There is no electrolyte in the probe (leak).
There may be a short-circuit in the connection.
B) The probe current is negative.
Connection problem to the anode circuit
(loose contact).
Deposits on the anode.
C) The sample temperature corresponds to the speci-
fications, but there is an error on the temperature.
There may be a short-circuit on the
temperature connection.
7-2
Other problems
PROBLEM: Inaccurate readings
Vista de pagina 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52 ... 55 56

Comentarios a estos manuales

Sin comentarios