Model 14 Oxygen Analyzer
SLOPE :
The flashing “CAL” message indicates that the
instrument is in calibration mode.
Wait for the
current to stabilize
(approximately 10 min.)
and press OK to validate the calibration. The
oxygen concentration value flashes for 3 seconds.
The analyzer displays the date of the last calibration
and the new slope calculated. This date can be
modified.
6-7
CONC. CALIB.
ZERO CALIB.
ZERO CALIB. SLOPE CALIB.
SLOPE CALIB.
CONC. CALIB.
ZERO CALIB.
SLOPE CALIB.
SLOPE CALIB.
1. Programming
ZERO :
Select chemical zero.
Select slope process.
2. Execution
ZERO :
Place the sensor in 99.999% Nitrogen gas.
Press ENTER, the “CAL” message flashes and
indicates the instrument is in calibration mode.
Wait for the current to stabilize (approximately
10 min.) and press OK to validate the calibration.
The oxygen concentration value flashes for 3
seconds.
The instrument will display zero.
6.2.4. Slope Calibration in Process with a Chemical Zero